Product Description
GF6019 DC Standard Power Source (Accuracy class 0.05/0.02)
Application It is suitable for power corporation measuring and
testing center, power supply company and power plants measuring
department, national levels measuring and testing institutions,
also suitable for railway, petroleum, chemical industry and large
industrial and mining enterprises, etc. Main functions and features
To calibrate DC voltmeter, ammeter, power meter and DC transmitter;
Automatic range, used for testing the digital meter; Manual range,
used for testing the indicator; To output the standard DC voltage,
DC current, DC power; To apply 5.6 inch TFT color LCD screen, all
the Chinese or English characters for display, operating easily; It
is built-in indicating meter and DC standard resource verification
procedures; Span a wide range, covering for instrument often dosage
limit; To use software calibration, stable and reliable; With RS232
interface, and computer software component semi-automatic
verification system, data management, inquiry and print the
inspection certificate and the inspection records; Technical Data
Voltage Output Range: 100 mV, 300 mV, 1 V, 3 V, 10 V, 100 V , 300 V
, 600 V , 1000 V Capacity 100 mV 40 mA 300 m V 100 mA 1 ~ 100 V 220
mA 300 V 80 mA 600 V 40 mA 1000 V 25 mA Current Output Range: 10
uA, 30 uA, 100 uA , 300 uA , 1 mA, 3 mA, 10 mA, 30 mA, 100 mA, 300
mA, 1 A, 3 A, 10 A, 30 A Capacity 10 uA ~ 1 A 10 V 3 A ~ 10 A 3 V
30 A 1.6 V Adjusting range and fine degrees 0 ~ 120% (1000 V
voltage, 30 A current except) Fineness 0.002% accuracy Voltage 1 V
and above range ± (0.012% RD + 0.008% RG); The rest of the range ±
(0.03% RD + 0.02% RG). All current range ± (0.03% RD + 0.02% RG)
(RD-reading RG-range) Full range of output stability A: for 0.75 V
voltage and the measuring range, all current range, 1 minutes full
range of output value ≤±0.01% RG ; B: for 1 V and above and voltage
range, 1 minutes full range of output value ≤±0.004% RG. Ripple
voltage noise caused by error A: ≤±0.01% RG, B: ≤±0.004% RG. Load
from no-load to full load cause of error A:≤±0.01% RG, B: ≤±0.004%
RG. Work power AC 220 V ±10%, frequency 50 Hz±5%. Use of the
environment temperature 20 ℃± 10 ℃, relative humidity ≤85% RH.
Shape dimension 440 mm x 360 mm x 160 mm, about 12 kg